COMPARISON BETWEEN PLACENTA ALPHA MICROGLOBULIN-I (Amnisure Test) AND STANDARD DIAGNOSTICS METHOD IN DETECTION OF RUPTURE OF FETAL MEMBRANES

  • AIPOH PETER OSHOKHAMELE National Postgraduate Medical College of Nigeria (NPMCN)

Abstract

Objective: To compare the diagnostic accuracy of placental alpha micro- globulin-1 assay with standard methods (Nitrazine stick test and clinical pooling of amniotic fluid in the posterior fornix) for detecting rupture of fetal membranes.

Study Design: Prospective simple comparative diagnostic study.
Setting: The study was conducted in Central Hospital,. Benin City, Nigeria

between February 2015 to June 2015.

Methods

A three hundred and twenty patients at gestational age of 24 – 42 weeks with history, signs and symptoms suggestive of drainage of liquor were recruited after informed consent. Each patient was tested for rupture of fetal membranes using placental alpha microglobulin-1 immunoassay, Nitrazine stick and visual inspection of postenior fornix for liquour drainage to determine the accuracy of PAMG-1 compared with standard diagnostic tests. Individual patient had a pre- test questionnaire. The data generated were analysed using basic descriptive statistics.
Main Outcome Measures-Placental alpha microglobulin -1 immunoassay and standard diagnostic methods (Nitrazine stick and pooling of liquour) method of rupture of fetal membranes.

Results- A total of 320 patients were recruited. 280 out of 320 patients 87.5% had actual ruptured membranes, whereas 40 patients 15.5% had intact membranes. Placental alpha microglobulin-1 immunoassay had sensitivity of 99.2% specificity of 89.36%, positive predictive value of 95.45%. In Negative predictive of value 42..9% (6/14). In contrast, the conventional standard diagnostic methods (pooling, and Nitrazine test) confirms ROM with a sensitivity of 81.2% (125/154), specificity of 50% (3/6) , positive predictive value of 97.7 %(125/128) and negative predictive value of 9.4%(3/32) .

Conclusion- The PAMG-1 immunoassay was a statistically better test in detecting ROM compared to the conventional standard diagnostic methods 94.8%(146/154) verus 81.2%(125/154) p< 0.000 Fisher’s test

Key words: Amnisure(R) Nitrazine, Pooling of liquor, rupture of fetal membraneaccuracy.